Error [I-01]: Batteries are getting low.

Solution: Replace batteries soon.

Error[I-02]: Meter Lifetime

Only -350 tests left before lifetime end. Contact your healthcare professional or Roche Diagnostics.

Error[I-03]: Wrong Code Chip

Solution: Wrong code chip is in the meter. Insert correct code chip.

Error [I-04]: Reminder in the Past

Solution: Only 99 days left until test strips expire. Make sure you have a new test strip vial availableError [I-06]: Test Strips ExpireSoon

Error [I-07]: Target/Trend only with INR

Solution: Target and Trend graphs are only available with unit INR. Change unit to INR to enable this function

Error [M-21]: Invalid Date

Solution: Date is invalid. Enter correct time and date.Error [M-22]: Test Strip Expired

Error [M-22): Test Strip Error

Solution: Test strip has expired. Use a valid test strip.

Error [M-23]: Test Strip Error

Solution: Barcode on test strip cannot be read. Repeat test with a new test stripError [M-25]: Invalid Code Chip

Error [M-24]: Test Strip Error

Solution: Meter does not recognize test strip. Repeat test with a new test strip.

Error [M-25]: Invalid Code Chip

Solution: Code chip is invalid. Replacewith a valid code chip.

Error [M-26]: Test Strip Error

Solution: Test strip removed during test. Repeat test with a new test strip

Error [M-27]: Cover Open

Solution: Test strip guide cover removed during test. Close cover and repeat test.

Error [M-28]: USB Cable / Light Interference

Solution: Connected USB cable or external light source interfering with test. Repeat test after disconnecting USB cable and protecting from light source.

Error [M-31]: Batteries Empty

Solution: Batteries are nearly empty. Replace batteries now

Error [M-32]: Ambient Temperature

Solution: Meter too cold or too warm. Power meter off. Repeat test after 30 minutes at room temperature.

Error [M-41]: No Blood Detected

Solution: No blood was detected. Repeat test, with a new test strip if necessary.

Error [M-42]: Sample Too Small

Solution: Not enough sample was applied. Repeat test with a new test strip.

Error [M-43]: Dosing Error

Solution: A dosing error was detected. Repeat test with a new test strip

Error [M-44]: Test Strip Error

Solution: Test strip failed internal control. Repeat test with a new test strip.

Error [M-71]: USB Error

Solution: A USB error has occurred. Power meter off and on and repeat connecting USB.

Error [M-72] [M-73]: Bluetooth Error

Solution: A Bluetooth error has occurred. Power meter off and on and repeat data transfer.

Error [E-33]: Meter Lifetime End

Solution: Maximum number of tests has been reached. Contact Reliaston for support.

Error [E-51]: Meter Error

Solution: Meter self-check detected a general error. Power meter off and on. If error persists, contact Roche Diagnostics.

Error [E-52]: Meter Error

Solution: Temperature or heater error. Please try again in 3-15 minutes. If error persists, contact Roche Diagnostics.

Error [E-61]: Meter Error

Solution: Meter is defective. No test is possible. Contact Roche Diagnostics.

Error [E-62]: Meter Error

Solution: Temperature or heater error. Power meter off and on. If error persists, contact Roche Diagnostics.

Error [E-63]: Meter Error

Solution: Hardware or electronic error. Power meter off and on. If error persists, contact Roche Diagnostics.

Error [E-64]: Meter Error

Solution: Test strip reader error. Power meter off and on. If error persists, contact Roche Diagnostics.

Error [E-65]: Meter Error

Solution: Code chip cannot be read. Replace with a different code chip.

Error [E-66]: Meter Error

Solution: A software error has occurred. Power meter off and on. If error persists, contact Roche Diagnostics

Error [E-67]: Meter Error

Solution: A memory error has occurred. Power meter off and on. If error persists, contact Roche Diagnostics.

Error [E-68]: Meter Power Failure

Solution: There was an unexpected interruption of the meter's power supply. Repeat your last procedure.